Filtered by vendor Philips
Subscribe
Total
100 CVE
CVE | Vendors | Products | Updated | CVSS v2 | CVSS v3 |
---|---|---|---|---|---|
CVE-2021-43550 | 1 Philips | 3 Efficia Cm, Efficia Cm Firmware, Patient Information Center Ix | 2022-01-12 | 3.3 LOW | 6.5 MEDIUM |
The use of a broken or risky cryptographic algorithm is an unnecessary risk that may result in the exposure of sensitive information, which affects the communications between Patient Information Center iX (PIC iX) Versions C.02 and C.03 and Efficia CM Series Revisions A.01 to C.0x and 4.0. | |||||
CVE-2021-43548 | 1 Philips | 1 Patient Information Center Ix | 2022-01-12 | 3.3 LOW | 6.5 MEDIUM |
Patient Information Center iX (PIC iX) Versions C.02 and C.03 receives input or data, but does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly. | |||||
CVE-2021-32993 | 1 Philips | 4 Intellibridge Ec40, Intellibridge Ec40 Firmware, Intellibridge Ec80 and 1 more | 2022-01-10 | 5.8 MEDIUM | 8.8 HIGH |
IntelliBridge EC 40 and 60 Hub (C.00.04 and prior) contains hard-coded credentials, such as a password or a cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data. | |||||
CVE-2021-33017 | 1 Philips | 4 Intellibridge Ec40, Intellibridge Ec40 Firmware, Intellibridge Ec80 and 1 more | 2022-01-10 | 5.8 MEDIUM | 8.8 HIGH |
The standard access path of the IntelliBridge EC 40 and 60 Hub (C.00.04 and prior) requires authentication, but the product has an alternate path or channel that does not require authentication. | |||||
CVE-2021-26248 | 1 Philips | 4 Mri 1.5t, Mri 1.5t Firmware, Mri 3t and 1 more | 2021-11-23 | 2.1 LOW | 5.5 MEDIUM |
Philips MRI 1.5T and MRI 3T Version 5.x.x assigns an owner who is outside the intended control sphere to a resource. | |||||
CVE-2020-16241 | 1 Philips | 2 Suresigns Vs4, Suresigns Vs4 Firmware | 2021-11-22 | 2.1 LOW | 2.1 LOW |
Philips SureSigns VS4, A.07.107 and prior. The software does not restrict or incorrectly restricts access to a resource from an unauthorized actor. | |||||
CVE-2021-39375 | 1 Philips | 1 Tasy Electronic Medical Record | 2021-09-14 | 6.5 MEDIUM | 8.8 HIGH |
Philips Healthcare Tasy Electronic Medical Record (EMR) 3.06 allows SQL injection via the WAdvancedFilter/getDimensionItemsByCode FilterValue parameter. | |||||
CVE-2021-39376 | 1 Philips | 1 Tasy Electronic Medical Record | 2021-08-31 | 6.5 MEDIUM | 8.8 HIGH |
Philips Healthcare Tasy Electronic Medical Record (EMR) 3.06 allows SQL injection via the CorCad_F2/executaConsultaEspecifico IE_CORPO_ASSIST or CD_USUARIO_CONVENIO parameter. | |||||
CVE-2018-10601 | 1 Philips | 36 Avalon Fetal\/maternal Monitors Fm20, Avalon Fetal\/maternal Monitors Fm20 Firmware, Avalon Fetal\/maternal Monitors Fm30 and 33 more | 2021-05-10 | 5.4 MEDIUM | 8.2 HIGH |
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow. | |||||
CVE-2018-10599 | 1 Philips | 36 Avalon Fetal\/maternal Monitors Fm20, Avalon Fetal\/maternal Monitors Fm20 Firmware, Avalon Fetal\/maternal Monitors Fm30 and 33 more | 2021-05-10 | 2.9 LOW | 5.3 MEDIUM |
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to read memory from an attacker-chosen device address within the same subnet. | |||||
CVE-2018-10597 | 1 Philips | 36 Avalon Fetal\/maternal Monitors Fm20, Avalon Fetal\/maternal Monitors Fm20 Firmware, Avalon Fetal\/maternal Monitors Fm30 and 33 more | 2021-05-10 | 5.4 MEDIUM | 8.3 HIGH |
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to access memory ("write-what-where") from an attacker-chosen device address within the same subnet. | |||||
CVE-2020-27298 | 1 Philips | 5 Coronary Tools, Dynamic Coronary Roadmap, Interventional Workspot and 2 more | 2021-02-02 | 3.3 LOW | 6.5 MEDIUM |
Philips Interventional Workspot (Release 1.3.2, 1.4.0, 1.4.1, 1.4.3, 1.4.5), Coronary Tools/Dynamic Coronary Roadmap/Stentboost Live (Release 1.0), ViewForum (Release 6.3V1L10). The software constructs all or part of an OS command using externally influenced input from an upstream component but does not neutralize or incorrectly neutralizes special elements that could modify the intended OS command when sent to a downstream component. | |||||
CVE-2019-10988 | 1 Philips | 2 Hdi 4000, Hdi 4000 Firmware | 2020-10-02 | 3.6 LOW | 3.4 LOW |
In Philips HDI 4000 Ultrasound Systems, all versions running on old, unsupported operating systems such as Windows 2000, the HDI 4000 Ultrasound System is built on an old operating system that is no longer supported. Thus, any unmitigated vulnerability in the old operating system could be exploited to affect this product. | |||||
CVE-2019-10968 | 1 Philips | 1 Zymed Holter 2010 | 2020-10-01 | 2.1 LOW | 4.4 MEDIUM |
Philips Holter 2010 Plus, all versions. A vulnerability has been identified that may allow system options that were not purchased to be enabled. | |||||
CVE-2018-8848 | 1 Philips | 1 E-alert Firmware | 2020-09-29 | 5.0 MEDIUM | 7.5 HIGH |
Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The software, upon installation, sets incorrect permissions for an object that exposes it to an unintended actor. | |||||
CVE-2020-16200 | 1 Philips | 1 Clinical Collaboration Platform | 2020-09-25 | 3.3 LOW | 6.5 MEDIUM |
Philips Clinical Collaboration Platform, Versions 12.2.1 and prior. The software does not properly control the allocation and maintenance of a limited resource, thereby enabling an attacker to influence the amount of resources consumed, eventually leading to the exhaustion of available resources. | |||||
CVE-2020-16198 | 1 Philips | 1 Clinical Collaboration Platform | 2020-09-25 | 5.8 MEDIUM | 6.3 MEDIUM |
Philips Clinical Collaboration Platform, Versions 12.2.1 and prior. When an attacker claims to have a given identity, the software does not prove or insufficiently proves the claim is correct. | |||||
CVE-2020-14525 | 1 Philips | 1 Clinical Collaboration Platform | 2020-09-25 | 2.7 LOW | 3.5 LOW |
Philips Clinical Collaboration Platform, Versions 12.2.1 and prior. The software does not neutralize or incorrectly neutralizes user-controllable input before it is placed in output used as a webpage that is served to other users. | |||||
CVE-2020-14506 | 1 Philips | 1 Clinical Collaboration Platform | 2020-09-25 | 4.3 MEDIUM | 4.3 MEDIUM |
Philips Clinical Collaboration Platform, Versions 12.2.1 and prior. The product receives input or data, but it does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly. | |||||
CVE-2018-17906 | 1 Philips | 2 Intellispace Pacs, Isite Pacs | 2020-09-18 | 3.3 LOW | 8.8 HIGH |
Philips iSite and IntelliSpace PACS, iSite PACS, all versions, and IntelliSpace PACS, all versions. Default credentials and no authentication within third party software may allow an attacker to compromise a component of the system. |