Filtered by vendor Hospira
Subscribe
Total
11 CVE
CVE | Vendors | Products | Updated | CVSS v2 | CVSS v3 |
---|---|---|---|---|---|
CVE-2014-5401 | 1 Hospira | 1 Mednet | 2019-10-09 | 10.0 HIGH | 9.8 CRITICAL |
Hospira MedNet software version 5.8 and prior uses vulnerable versions of the JBoss Enterprise Application Platform software that may allow unauthenticated users to execute arbitrary code on the target system. Hospira has developed a new version of the MedNet software, MedNet 6.1. Existing versions of MedNet can be upgraded to MedNet 6.1. | |||||
CVE-2015-3459 | 1 Hospira | 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware | 2017-01-03 | 10.0 HIGH | N/A |
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands. | |||||
CVE-2015-3955 | 1 Hospira | 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware | 2016-12-05 | 10.0 HIGH | N/A |
Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors. | |||||
CVE-2015-3958 | 1 Hospira | 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware | 2016-12-05 | 7.8 HIGH | N/A |
Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to cause a denial of service (forced manual reboot) via a flood of TCP packets. | |||||
CVE-2015-3957 | 1 Hospira | 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware | 2016-12-05 | 4.6 MEDIUM | N/A |
Hospira LifeCare PCA Infusion System before 7.0 stores private keys and certificates, which has unspecified impact and attack vectors. | |||||
CVE-2015-7909 | 1 Hospira | 2 Communication Engine, Lifecare Pca Infusion System | 2016-02-09 | 7.5 HIGH | 7.3 HIGH |
Stack-based buffer overflow in Hospira Communication Engine (CE) before 1.2 in LifeCare PCA Infusion System 5.07, Plum A+ Infusion System 13.40, and Plum A+3 Infusion System 13.40 allows remote attackers to cause a denial of service or possibly have unspecified other impact via traffic on TCP port 5000. | |||||
CVE-2014-5405 | 1 Hospira | 1 Mednet | 2015-07-24 | 9.0 HIGH | N/A |
Hospira MedNet before 6.1 uses a hardcoded cleartext password to control SQL database authorization, which allows remote authenticated users to bypass intended access restrictions by leveraging knowledge of this password. | |||||
CVE-2015-1011 | 1 Hospira | 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware | 2015-07-08 | 5.0 MEDIUM | N/A |
Hospira LifeCare PCA Infusion System before 7.0 has hardcoded credentials, which makes it easier for remote attackers to obtain access via unspecified vectors. | |||||
CVE-2014-5406 | 1 Hospira | 3 Lifecare Pca3, Lifecare Pca5, Lifecare Pcainfusion Firmware | 2015-07-08 | 9.3 HIGH | N/A |
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459. | |||||
CVE-2014-5400 | 1 Hospira | 1 Mednet | 2015-04-03 | 2.1 LOW | N/A |
The installation component in Hospira MedNet before 6.1 places cleartext credentials in configuration files, which allows local users to obtain sensitive information by reading a file. | |||||
CVE-2014-5403 | 1 Hospira | 1 Mednet | 2015-04-03 | 5.0 MEDIUM | N/A |
Hospira MedNet before 6.1 uses hardcoded cryptographic keys for protection of data transmission from infusion pumps, which allows remote attackers to obtain sensitive information by sniffing the network. |