Vulnerabilities (CVE)

Join the Common Vulnerabilities and Exposures (CVE) community and start to get notified about new vulnerabilities.

Filtered by vendor Baxter Subscribe
Filtered by product Prismaflex Firmware
Total 3 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2020-12037 1 Baxter 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more 2020-07-14 5.0 MEDIUM 7.5 HIGH
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
CVE-2020-12036 1 Baxter 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more 2020-07-14 5.0 MEDIUM 7.5 HIGH
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
CVE-2020-12035 1 Baxter 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more 2020-07-14 3.6 LOW 4.9 MEDIUM
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.