Total
3 CVE
CVE | Vendors | Products | Updated | CVSS v2 | CVSS v3 |
---|---|---|---|---|---|
CVE-2020-12037 | 1 Baxter | 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more | 2020-07-14 | 5.0 MEDIUM | 7.5 HIGH |
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device. | |||||
CVE-2020-12036 | 1 Baxter | 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more | 2020-07-14 | 5.0 MEDIUM | 7.5 HIGH |
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device. | |||||
CVE-2020-12035 | 1 Baxter | 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more | 2020-07-14 | 3.6 LOW | 4.9 MEDIUM |
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration. |