CVE-2021-23195 | 1 Fresenius-kabi | 8 Agilia Connect, Agilia Connect Firmware, Agilia Partner Maintenance Software and 5 more | 2022-10-24 | 5.0 MEDIUM | 5.3 MEDIUM |
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 has the option for automated indexing (directory listing) activated. When accessing a directory, a web server delivers its entire content in HTML form. If an index file does not exist and directory listing is enabled, all content of the directory will be displayed, allowing an attacker to identify and access files on the server. | |||||
CVE-2021-23196 | 1 Fresenius-kabi | 8 Agilia Connect, Agilia Connect Firmware, Agilia Partner Maintenance Software and 5 more | 2022-08-30 | 7.5 HIGH | 9.8 CRITICAL |
The web application on Agilia Link+ version 3.0 implements authentication and session management mechanisms exclusively on the client-side and does not protect authentication attributes sufficiently. | |||||
CVE-2021-23207 | 1 Fresenius-kabi | 7 Agilia Connect, Agilia Partner Maintenance Software, Link\+ Agilia and 4 more | 2022-08-30 | 2.1 LOW | 5.5 MEDIUM |
An attacker with physical access to the host can extract the secrets from the registry and create valid JWT tokens for the Fresenius Kabi Vigilant MasterMed version 2.0.1.3 application and impersonate arbitrary users. An attacker could manipulate RabbitMQ queues and messages by impersonating users. | |||||
CVE-2021-43355 | 1 Fresenius-kabi | 8 Agilia Connect, Agilia Connect Firmware, Agilia Partner Maintenance Software and 5 more | 2022-08-08 | 7.5 HIGH | 9.8 CRITICAL |
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support or block JavaScript or intentionally bypass the client-side checks. An attacker with knowledge of the service user could circumvent the client-side control and login with service privileges. | |||||
CVE-2021-23236 | 1 Fresenius-kabi | 8 Agilia Connect, Agilia Connect Firmware, Agilia Partner Maintenance Software and 5 more | 2022-01-28 | 7.8 HIGH | 7.5 HIGH |
Requests may be used to interrupt the normal operation of the device. When exploited, Fresenius Kabi Agilia Link+ version 3.0 must be rebooted via a hard reset triggered by pressing a button on the rack system. | |||||
CVE-2021-23233 | 1 Fresenius-kabi | 8 Agilia Connect, Agilia Connect Firmware, Agilia Partner Maintenance Software and 5 more | 2022-01-28 | 7.5 HIGH | 9.8 CRITICAL |
Sensitive endpoints in Fresenius Kabi Agilia Link+ v3.0 and prior can be accessed without any authentication information such as the session cookie. An attacker can send requests to sensitive endpoints as an unauthenticated user to perform critical actions or modify critical configuration parameters. | |||||
CVE-2021-44464 | 1 Fresenius-kabi | 8 Agilia Connect, Agilia Connect Firmware, Agilia Partner Maintenance Software and 5 more | 2022-01-28 | 6.5 MEDIUM | 8.8 HIGH |
Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 contains service credentials likely to be common across all instances. An attacker in possession of the password may gain privileges on all installations of this software. | |||||
CVE-2021-33848 | 1 Fresenius-kabi | 8 Agilia Connect, Agilia Connect Firmware, Agilia Partner Maintenance Software and 5 more | 2022-01-28 | 4.3 MEDIUM | 6.1 MEDIUM |
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 is vulnerable to reflected cross-site scripting attacks. An attacker could inject JavaScript in a GET parameter of HTTP requests and perform unauthorized actions such as stealing internal information and performing actions in context of an authenticated user. | |||||
CVE-2021-33846 | 1 Fresenius-kabi | 8 Agilia Connect, Agilia Connect Firmware, Agilia Partner Maintenance Software and 5 more | 2022-01-28 | 6.5 MEDIUM | 7.2 HIGH |
Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 issues authentication tokens to authenticated users that are signed with a symmetric encryption key. An attacker in possession of the key can issue valid JWTs and impersonate arbitrary users. | |||||
CVE-2021-41835 | 1 Fresenius-kabi | 7 Agilia Connect, Agilia Partner Maintenance Software, Link\+ Agilia and 4 more | 2022-01-27 | 5.0 MEDIUM | 7.5 HIGH |
Fresenius Kabi Agilia Link + version 3.0 does not enforce transport layer encryption. Therefore, transmitted data may be sent in cleartext. Transport layer encryption is offered on Port TCP/443, but the affected service does not perform an automated redirect from the unencrypted service on Port TCP/80 to the encrypted service. |