Fresenius Kabi Agilia Link + version 3.0 does not enforce transport layer encryption. Therefore, transmitted data may be sent in cleartext. Transport layer encryption is offered on Port TCP/443, but the affected service does not perform an automated redirect from the unencrypted service on Port TCP/80 to the encrypted service.
References
Link | Resource |
---|---|
https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01 | Third Party Advisory US Government Resource |
Configurations
Configuration 1 (hide)
|
Configuration 2 (hide)
AND |
|
Configuration 3 (hide)
AND |
|
Information
Published : 2022-01-21 11:15
Updated : 2022-01-27 11:26
NVD link : CVE-2021-41835
Mitre link : CVE-2021-41835
JSON object : View
CWE
CWE-319
Cleartext Transmission of Sensitive Information
Products Affected
fresenius-kabi
- link\+_agilia_firmware
- agilia_connect
- vigilant_mastermed
- link\+_agilia
- vigilant_centerium
- vigilant_insight
- agilia_partner_maintenance_software