Filtered by vendor Baxter
Subscribe
Total
25 CVE
CVE | Vendors | Products | Updated | CVSS v2 | CVSS v3 |
---|---|---|---|---|---|
CVE-2020-12012 | 1 Baxter | 4 Em1200, Em1200 Firmware, Em2400 and 1 more | 2020-07-07 | 3.6 LOW | 6.1 MEDIUM |
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI. | |||||
CVE-2014-5432 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2019-10-09 | 7.5 HIGH | 9.8 CRITICAL |
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes. | |||||
CVE-2014-5434 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2019-10-09 | 5.0 MEDIUM | 9.8 CRITICAL |
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes. | |||||
CVE-2014-5433 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2019-10-09 | 7.5 HIGH | 9.8 CRITICAL |
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes. | |||||
CVE-2014-5431 | 1 Baxter | 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module | 2019-10-09 | 4.6 MEDIUM | 6.8 MEDIUM |
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes. |