CVE-2014-5433

An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
References
Link Resource
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 Mitigation US Government Resource Third Party Advisory
Advertisement

NeevaHost hosting service

Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:*
OR cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:*
cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*

Information

Published : 2019-03-26 09:29

Updated : 2019-10-09 16:11


NVD link : CVE-2014-5433