An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
References
Link | Resource |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 | Mitigation US Government Resource Third Party Advisory |
Configurations
Configuration 1 (hide)
AND |
|
Information
Published : 2019-03-26 09:29
Updated : 2019-10-09 16:11
NVD link : CVE-2014-5433
Mitre link : CVE-2014-5433
JSON object : View
CWE
CWE-255
Credentials Management Errors
Products Affected
baxter
- wireless_battery_module
- sigma_spectrum_infusion_system
- sigma_spectrum_infusion_system_firmware