Vulnerabilities (CVE)

Join the Common Vulnerabilities and Exposures (CVE) community and start to get notified about new vulnerabilities.

Filtered by vendor Niscomed Subscribe
Filtered by product M1000 Multipara Patient Monitor Firmware
Total 4 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2020-15482 1 Niscomed 2 M1000 Multipara Patient Monitor, M1000 Multipara Patient Monitor Firmware 2021-07-21 7.2 HIGH 7.8 HIGH
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The device enables an unencrypted TELNET service by default, with a blank password for the admin account. This allows an attacker to gain root access to the device over the local network.
CVE-2020-15483 1 Niscomed 2 M1000 Multipara Patient Monitor, M1000 Multipara Patient Monitor Firmware 2021-07-21 7.2 HIGH 6.8 MEDIUM
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The physical UART debug port provides a shell, without requiring a password, with complete access.
CVE-2020-15485 1 Niscomed 2 M1000 Multipara Patient Monitor, M1000 Multipara Patient Monitor Firmware 2021-07-15 2.1 LOW 5.5 MEDIUM
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The onboard Flash memory stores data in cleartext, without integrity protection against tampering.
CVE-2020-15484 1 Niscomed 2 M1000 Multipara Patient Monitor, M1000 Multipara Patient Monitor Firmware 2020-08-31 5.0 MEDIUM 7.5 HIGH
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The internal storage of the underlying Linux system stores data in cleartext, without integrity protection against tampering.