Vulnerabilities (CVE)

Join the Common Vulnerabilities and Exposures (CVE) community and start to get notified about new vulnerabilities.

Filtered by vendor Medtronicdiabetes Subscribe
Filtered by product 508 Minimed Insulin Pump Firmware
Total 1 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2018-14781 1 Medtronicdiabetes 18 508 Minimed Insulin Pump, 508 Minimed Insulin Pump Firmware, 522 Paradigm Real-time and 15 more 2019-10-09 2.9 LOW 5.3 MEDIUM
Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.