CVE-2022-38100

The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
References
Link Resource
https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 Mitigation Third Party Advisory US Government Resource
Advertisement

NeevaHost hosting service

Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*

Information

Published : 2022-09-13 08:15

Updated : 2022-09-14 15:48


NVD link : CVE-2022-38100

Mitre link : CVE-2022-38100


JSON object : View

CWE
CWE-400

Uncontrolled Resource Consumption

Advertisement

dedicated server usa

Products Affected

contechealth

  • cms8000
  • cms8000_firmware