A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and takes extensive time for the GUI to connect to the TCP socket, allowing the connection to be hijacked by an external attacker.
References
Configurations
Configuration 1 (hide)
AND |
|
Information
Published : 2021-08-02 06:15
Updated : 2021-08-10 12:44
NVD link : CVE-2021-37166
Mitre link : CVE-2021-37166
JSON object : View
CWE
CWE-120
Buffer Copy without Checking Size of Input ('Classic Buffer Overflow')
Products Affected
swisslog-healthcare
- hmi-3_control_panel_firmware
- hmi-3_control_panel