A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote code execution.
References
Configurations
Configuration 1 (hide)
AND |
|
Information
Published : 2021-08-02 06:15
Updated : 2021-08-10 12:06
NVD link : CVE-2021-37162
Mitre link : CVE-2021-37162
JSON object : View
CWE
CWE-120
Buffer Copy without Checking Size of Input ('Classic Buffer Overflow')
Products Affected
swisslog-healthcare
- hmi-3_control_panel_firmware
- hmi-3_control_panel