An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
References
Link | Resource |
---|---|
https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/ | Exploit Third Party Advisory |
https://www.bbraunusa.com/en.htm | Broken Link |
Configurations
Configuration 1 (hide)
AND |
|
Information
Published : 2021-08-25 05:15
Updated : 2022-07-12 10:42
NVD link : CVE-2021-33885
Mitre link : CVE-2021-33885
JSON object : View
CWE
CWE-347
Improper Verification of Cryptographic Signature
Products Affected
bbraun
- infusomat_large_volume_pump_871305u
- spacestation_8713142u
- spacecom2