A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
References
Link | Resource |
---|---|
https://www.siemens-healthineers.com/support-documentation/security-advisory | Vendor Advisory |
Configurations
Configuration 1 (hide)
AND |
|
Information
Published : 2020-10-13 09:15
Updated : 2020-10-29 13:12
NVD link : CVE-2020-7590
Mitre link : CVE-2020-7590
JSON object : View
CWE
CWE-259
Use of Hard-coded Password
Products Affected
siemens
- dca_vantage_analyzer
- dca_vantage_analyzer_firmware