Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior. The software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart.
References
Link | Resource |
---|---|
https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01 | Third Party Advisory US Government Resource |
Configurations
Configuration 1 (hide)
|
Information
Published : 2020-09-11 07:15
Updated : 2020-09-15 11:18
NVD link : CVE-2020-16224
Mitre link : CVE-2020-16224
JSON object : View
CWE
CWE-130
Improper Handling of Length Parameter Inconsistency
Products Affected
philips
- patient_information_center_ix