Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
References
Link | Resource |
---|---|
https://www.us-cert.gov/ics/advisories/icsma-20-170-04 | Third Party Advisory US Government Resource |
Configurations
Configuration 1 (hide)
AND |
|
Information
Published : 2020-06-29 07:15
Updated : 2020-07-09 10:58
NVD link : CVE-2020-12040
Mitre link : CVE-2020-12040
JSON object : View
CWE
CWE-319
Cleartext Transmission of Sensitive Information
Products Affected
baxter
- sigma_spectrum_infusion_system
- sigma_spectrum_infusion_system_firmware