Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.
References
Link | Resource |
---|---|
https://www.us-cert.gov/ics/advisories/icsma-20-170-04 | Third Party Advisory US Government Resource |
Configurations
Configuration 1 (hide)
AND |
|
Information
Published : 2020-06-29 07:15
Updated : 2020-07-09 10:52
NVD link : CVE-2020-12039
Mitre link : CVE-2020-12039
JSON object : View
CWE
CWE-798
Use of Hard-coded Credentials
Products Affected
baxter
- sigma_spectrum_infusion_system
- sigma_spectrum_infusion_system_firmware