CVE-2018-8842

Philips e-Alert Unit (non-medical device), Version R2.1 and prior. The software transmits sensitive or security-critical data in cleartext in a communication channel that can be sniffed by unauthorized actors. The Philips e-Alert communication channel is not encrypted which could therefore lead to disclosure of personal contact information and application login credentials from within the same subnet.
References
Link Resource
https://www.usa.philips.com/healthcare/about/customer-support/product-security Vendor Advisory
https://ics-cert.us-cert.gov/advisories/ICSA-18-242-01 Mitigation Third Party Advisory US Government Resource
http://www.securityfocus.com/bid/105194 Third Party Advisory VDB Entry
Advertisement

NeevaHost hosting service

Configurations

Configuration 1 (hide)

cpe:2.3:o:philips:e-alert_firmware:*:*:*:*:*:*:*:*

Information

Published : 2018-09-26 12:29

Updated : 2019-10-09 16:42


NVD link : CVE-2018-8842

Mitre link : CVE-2018-8842


JSON object : View

CWE
CWE-319

Cleartext Transmission of Sensitive Information

Advertisement

dedicated server usa

Products Affected

philips

  • e-alert_firmware