Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
References
Link | Resource |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01 | Third Party Advisory US Government Resource |
http://www.securityfocus.com/bid/100523 | Third Party Advisory VDB Entry |
Configurations
Configuration 1 (hide)
AND |
|
Configuration 2 (hide)
AND |
|
Configuration 3 (hide)
AND |
|
Configuration 4 (hide)
AND |
|
Information
Published : 2018-04-25 06:29
Updated : 2019-10-09 16:23
NVD link : CVE-2017-12716
Mitre link : CVE-2017-12716
JSON object : View
CWE
CWE-319
Cleartext Transmission of Sensitive Information
Products Affected
abbott
- anthem
- accent_mri_firmware
- accent_st_firmware
- accent_mri
- accent_st
- accent
- accent_firmware
- anthem_firmware