CVE-2017-12716

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
References
Link Resource
https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01 Third Party Advisory US Government Resource
http://www.securityfocus.com/bid/100523 Third Party Advisory VDB Entry
Advertisement

NeevaHost hosting service

Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:abbott:accent_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:abbott:accent:-:*:*:*:*:*:*:*

Configuration 2 (hide)

AND
cpe:2.3:o:abbott:anthem_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:abbott:anthem:-:*:*:*:*:*:*:*

Configuration 3 (hide)

AND
cpe:2.3:o:abbott:accent_mri_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:abbott:accent_mri:-:*:*:*:*:*:*:*

Configuration 4 (hide)

AND
cpe:2.3:o:abbott:accent_st_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:abbott:accent_st:-:*:*:*:*:*:*:*

Information

Published : 2018-04-25 06:29

Updated : 2019-10-09 16:23


NVD link : CVE-2017-12716

Mitre link : CVE-2017-12716


JSON object : View

CWE
CWE-319

Cleartext Transmission of Sensitive Information

Advertisement

dedicated server usa

Products Affected

abbott

  • anthem
  • accent_mri_firmware
  • accent_st_firmware
  • accent_mri
  • accent_st
  • accent
  • accent_firmware
  • anthem_firmware